Some operators like to turn the "current adjustment" potentiometer to the left, so that the no-load voltage of the power supply cannot be adjusted. This shows his lack of substantial understanding of "current regulation". Because the power supply consumes a little current even if it is no-load, and if you turn off the "current regulation" to zero, even a small current will not be released, and of course the no-load voltage will not rise. Therefore, "current regulation" is generally not adjusted to zero (the above situation will not occur if it is adjusted to the right to about a quarter of a turn).


Opći način rada: Kada se koristi izlazni napon reguliranog izvora, gumb za podešavanje struje treba okrenuti u smjeru kazaljke na satu do kraja i zadržati. Podesite gumb za podešavanje napona kako biste kontrolirali vrijednost izlaznog istosmjernog napona. Kada dajete struju kao stalni izvor struje, okrenite gumb za podešavanje napona u smjeru kazaljke na satu do kraja i zadržite ga. Podesite gumb za podešavanje struje kako biste kontrolirali vrijednost izlazne istosmjerne struje.